Edwards Logo

Vacuum Measurement and control systems - Edwards (formely known as BOC Edwards) vacuum guages


Edwards (formerly BOC Edwards) ATC Active Thermocouple Gauge

65 to 1.3 x 10-3 mbar


The Edwards (formerly known as BOC Edwards) active Thermocouple (ATC-E) gauges mount directly on the thermocouple tube to form a compact stand alone transducer. The ATC-E electronics module drives both medium and low pressure gauge tubes. These gauges offer a cost effective measuring solution for higher pressures.

Edwards (formerly BOC Edwards) ATC Active Thermocouple Gauge

Features and Benefits

  • Drive electronics mount directly on the gauge tube
    Simplifies system design
    Saves valuable rack space
  • Wide range, regulated, internal power supply
    Runs from standard d.c power supplies
    +13.5 to +36 V
    Tolerant to voltage fluctuations
  • Standard analogue output 0 to +10 v d.c. And gauge identifier
    Easy to interface with a computer or PLC
    Fault operation indication
  • Adjustable set point with vacuum status LED
    Use for process control and interlocking
    Includes a digital vacuum status signal
    Set-point level ready visually
  • Low output impedance and integral Faraday shield
    Provides high level of noise immunity
    Permits long cable runs (up tp 100 metres)
  • Choice of medium and low pressure gauge tubes
    ATC-D covers from 50 to 5 x 10-2 Torr
  • ATC-M covers from 1 to 10-3 Torr
Edwards (formerly BOC Edwards) ATC Active Thermocouple Gauge

Technical specification

ATC-E electronics module


Power Supply: +13.5 to +36 V d.c. (max 1 volt ripple)
Power consumption: 0.54 W (max)
Output signal:

Operating: 2 to 10 V d.c.
Fault Output: 0 to 2 V d.c. 10 to 13.5V d.c.
Output Impedance: 0.1 Ohm
Minimum Load: 10 k Ohms
Tube type selection
Adjustments: Adjust set point visually via potentiometer
Set point: Open collector transistor
Range of set point: 2% to 85% full scale voltage
Fixed hysteresis: 12% full scale voltage
Level settings: ± 2% full scale voltage
Rating: 40 V d.c. 100mA max
Temperature range:
Operating temperature: +5 to +60 ºC
Storage temperature: 0 to +70 ºC
Weight: 110g
External interface connector: 8 way FCC68(RJ45) socket
Standards
Electronic design: BS EN 61010-1
Eletromagnetic compatibility: EN 61326 Industrial location Class B emissions
Flame resistant casing: UL 94 (V0)
ATC-D, ATC-M gauge tubes
Pressure Range
ATC-D
50 to 5 x 10-2 Torr
(65 to 6.5 x 10-2 mbar)
ATC-M
1 to 1 x 10-3 Torr
(1.3 to 1.3 x 10-3 Torr)
Maximum over pressure:
ATC-D: 10 bar absolute
ATC-M: 3.4 bar absolute
Weight: 55g
Internal Volume
ATC-D/ATC-M: 1 cm3/8cm3
Vacuum fitting 1/8” NPT
Pin allocation:
1: power supply positive
2: power supply common
3: Gauge output
4: Gauge identification
5: Signal common
6: Set point output
7: Not connected
8: Not connected

Edwards (formerly BOC Edwards) ATC Active Thermocouple Gauge

Ordering information

Product Description Ordering Number
ATC-E Electronics module D35108000
ATC-D-1/6 inch NPT gauge tube D35512000
ATC-M-1/8 inch NPT gauge tube D35513000
Accessories and Spares
1/8 inch NPT to NW10 adaptor C10501072
Surge protector D40006000
Cables
0.5m active gauge cable D40001005
1m active gauge cable D40001010
3m active gauge cable D40001030
5m active gauge cable D40001050
10m active gauge cable D40001100
15m active gauge cable D40001150
25m active gauge cable D40001250
15m active gauge cable (24 AWG) D40005150
25m active gauge cable (24 AWG) D40005300

Non-standard lengths and screened cables available on request